# Start of file 60383_artiq_log.txt # First point: 60383.269973 2024-03-14T06:28:45.629000+00:00 # Last point: 60383.716641 2024-03-14T17:11:57.740000+00:00 # Num rows: 23 # File updated: 2024-03-15T01:10:31.925289 # Columns: MJD DATE RID PROGRAM MESSAGE 60383.269972558 2024-03-14T06:28:45.629000+00:00 37419 StateMachineSupervisor Start Clockrun 60383.269998704 2024-03-14T06:28:47.888000+00:00 37421 switchBeam Beam: B 60383.270010833 2024-03-14T06:28:48.936000+00:00 37422 Lock_to_C124 t_probe [us] = 136000, n_shots = 2, g1 = 0.3, T_drift_s = 600.0, N_drift = 450, pmt_thr = 7, T_p0avg_s = 900.0, N_p0avg = 225, (C1, C2, C4) att [dB] = (15.0, 15.5, 15.5), LCplus [V] = 1.8443, LCminus [V] = 3.2549 60383.603593287 2024-03-14T14:29:10.460000+00:00 37419 StateMachineSupervisor Ion Dark 60383.603599317 2024-03-14T14:29:10.981000+00:00 37419 StateMachineSupervisor Recover Ion 60383.603609283 2024-03-14T14:29:11.842000+00:00 37423 switchBeam Beam: ABC 60383.603644329 2024-03-14T14:29:14.870000+00:00 37426 SetCoolingAOM mode : recovery, AOM_f (MHz) : 180.0, ATT_Main (dB) = 5.0, ATT_SP (dB) = 12.0. 60383.603739826 2024-03-14T14:29:23.121000+00:00 37427 SetCoolingAOM mode : clockrun, AOM_f (MHz) : 208.0, ATT_Main (dB) = 12.0, ATT_SP (dB) = 22.0. 60383.603759433 2024-03-14T14:29:24.815000+00:00 37429 switchBeam Beam: B 60383.603783819 2024-03-14T14:29:26.922000+00:00 37430 MeasureFluorescenceLevels Bright : 37.2 +- 5.5, Dark : 0.2 +- 0.5, Threshold: int( ceil( 3.0 ) ) = 4, N_meas = 100, t_gate = 5.0 ms 60383.603797361 2024-03-14T14:29:28.092000+00:00 37419 StateMachineSupervisor Ion Bright, Start Clockrun 60383.603833785 2024-03-14T14:29:31.239000+00:00 37433 switchBeam Beam: B 60383.603844665 2024-03-14T14:29:32.179000+00:00 37434 Lock_to_C124 t_probe [us] = 136000, n_shots = 2, g1 = 0.3, T_drift_s = 600.0, N_drift = 450, pmt_thr = 7, T_p0avg_s = 900.0, N_p0avg = 225, (C1, C2, C4) att [dB] = (15.0, 15.5, 15.5), LCplus [V] = 1.8443, LCminus [V] = 3.2549 60383.706005729 2024-03-14T16:56:38.895000+00:00 37436 SetCoolingAOM mode : recovery, AOM_f (MHz) : 180.0, ATT_Main (dB) = 5.0, ATT_SP (dB) = 12.0. 60383.706686748 2024-03-14T16:57:37.735000+00:00 37437 switchBeam Beam: ABC 60383.707443576 2024-03-14T16:58:43.125000+00:00 37438 SetCoolingAOM mode : clockrun, AOM_f (MHz) : 208.0, ATT_Main (dB) = 12.0, ATT_SP (dB) = 22.0. 60383.707496597 2024-03-14T16:58:47.706000+00:00 37439 switchBeam Beam: B 60383.711835891 2024-03-14T17:05:02.621000+00:00 37440 MicroMotionMeasurement Beam: A, N_photons: 100000, N_bins: 25, f_mod [MHz]: 14.424000, duration [s]: 16.403 60383.712044618 2024-03-14T17:05:20.655000+00:00 37441 MicroMotionMeasurement Beam: B, N_photons: 100000, N_bins: 25, f_mod [MHz]: 14.424000, duration [s]: 18.024 60383.712208148 2024-03-14T17:05:34.784000+00:00 37442 MicroMotionMeasurement Beam: C, N_photons: 100000, N_bins: 25, f_mod [MHz]: 14.424000, duration [s]: 14.119 60383.712778275 2024-03-14T17:06:24.043000+00:00 37443 switchBeam Beam: B 60383.716531667 2024-03-14T17:11:48.336000+00:00 37444 CoolingDetuningScan From [AOM MHz]: 175.000, To [AOM MHz]: 220.000, num_pts: 100, t_gate (ms) = 1000.0 60383.716640509 2024-03-14T17:11:57.740000+00:00 37445 MeasureFluorescenceLevels Bright : 36.9 +- 5.7, Dark : 0.3 +- 0.5, Threshold: int( ceil( 3.1 ) ) = 4, N_meas = 100, t_gate = 5.0 ms # End of file 60383_artiq_log.txt